FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
121
rates. Typically this is used to take images at a high rate while
compound is being added, and shortly thereafter, to best capture
kinetic effects, then images can be taken at a less frequent rate as cells
settle and less change occurs.
The table below describes the options available in the Read process
page:
Note:
The read rate you configure for the Read process is the same
for all the read modes you have enabled. For example, if you set the
Read Interval to 1 s, and have two read modes, two images are taken
every second—one for each read mode.
Item
Description
First Interval
Configure the first series of reads taken during the
Read
process.
Read Interval
—Enter time (in seconds) between reads. The
same interval is used for each read mode (for example
,
the
instrument takes images twice as fast if a second read mode
is enabled).
Note:
Time should be no less than exposure time
plus read-out time (see
for more information). If it is less, then the
experiment speed is determined by the capability of
the computer.
Number of Reads
—Enter the total number of reads to be
taken in the first interval.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...