Calibration and Signal Test
150
0112-0109 H
7.
Enter the
Plate Specifications
:
Enter a
Plate Name
.
Select
Read Plate
as the
Plate Type
.
Type in the plate dimensions.
8.
Click
Finish
to create a new read plate with mask.
Plate mask definition is complete when the green
Assay Finished
(Unlocked)
light is illuminated.
Flat Fielding
Microplate illumination during a fluorescence assay does not result in
perfectly uniform excitation distribution on the microplate surface. In
the FLIPR
®
Tetra System, light distribution is more intense in the
middle of the microplate than the edges, therefore a flat-fielding
algorithm is applied to the microplate image to compensate for
variations in light intensity across the microplate.
During luminescence assays LEDs are not used. Flat-field correction are
preset during installation and do not require any user adjustment.
Recalibrating the Optics
FLIPR
®
Tetra System identifies the serial number, wavelength range and
position of each LED bank present in the instrument. This information,
in addition to the wavelength range for the emission filters, is stored
with the system’s calibration files. Any time LEDs or emission filters are
changed in the FLIPR
®
Tetra System, you need to recalibrate the Flat-
Field Calibration files for the new LED/Filter pairs. The following
procedure steps you through the process of recalibrating the system
optics.
If you have any questions or concerns regarding the recalibration
procedure, please contact Customer Service at +1-800-635-5577.
To recalibrate the optics:
1.
Select the
Calibration
command from the
Instrument
menu,
or click on the
Calibration
button.
2.
The
ScreenWorks Calibration
dialog appears. This dialog
includes a list of LED/Filter wavelength combinations and a list
of whether the calibration is
Done
or
Not Done
for Flat-Field
Calibration for the selected plate format.
Note:
Due to the improved light tightness of the new FLIPR
®
Tetra
Systems, the background correction calibration available in the
previous versions of this software (used with black bottom plates) is
no longer required.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...