Code no. 11
Message:
Ini-file has been updated. The versions are now matching.
Cause:
The software configuration has been upgraded
Action:
Press OK to confirm
Code no. 12
Message:
Calibration values are missing
Cause:
The system’s calibration data is missing.
Action:
If the Calibration Tool is available, a calibration of the system can be performed. For more
information, consult the chapter ‘Calibration’.
Contact Megger technical assistance if the problem persists.
Code no. 13
Message:
Free disk space has reached the warning level. Please delete old log-files.
Cause:
Too many reports are stored in the BALTO System.
Action:
Clear the stored report files by pressing ‘Delete’ in the ‘Report’ dialog.
This message does not require immediate action.
Contact Megger technical assistance if the problem persists.
Code no. 14
Message:
** Free disk space has reached the CRITICAL LEVEL ** Please delete old log files as soon as
possible.
Cause:
Too many reports are stored in the BALTO System.
Action:
Clear the stored report files by pressing ‘Delete’ in the ‘Report’ dialog.
This message requires immediate action.
Contact Megger technical assistance if the problem persists.
Code no. 15
(Reserved for internal use.)
Code no. 16
(Reserved for internal use.)
Code no. 17
(Reserved for internal use.)
Code no. 18
Message:
Files copied.
Cause:
The operator has inserted a USB stick. All report files have been transferred to it.
Action:
Press OK to confirm.
Code no. 19
Message:
USB Stick removed.
Cause:
The USB stick has been disconnected.
Action:
Press OK to confirm.
Code no. 20
Message:
This Control Unit is configured for … Power Units, but you are using … Power Unit(s). Please
check the connections to the HS DC CBR and disconnect the Power Cables from unused Power
Units!
Cause:
The number of connected DC Current Generators does not match the default factory
configuration.
Action:
If the different number of DC Current Generators is intentional, press OK to confirm.
Ensure that unused DC Current Generators are disconnected from the High-speed DC Circuit
Breaker.
Code no. 21
Message:
You have DISCONNECTED some Power Units. Please do not forget to DISCONNECT the
corresponding Power Cables at the HS DC CBR side.
Cause:
A DC Current Generator has been disconnected from the system.
Action:
Ensure that any disconnected DC Current Generator is also disconnected from the High-speed
DC Circuit Breaker.
www.megger.com
161
Troubleshooting.