12.3 Message codes.
Messages, warnings and errors displayed in the calibration program are assigned a unique code. A list of message
codes is included below.
12.3.1 General messages.
Message no.
Cause:
Action:
1:
Battery voltage too
low (critical).
Battery voltage < 13
VDC
Please wait for the battery to recharge before performing
the calibration step.
The battery indicator must be green to continue.
2:
Charging Current too
high
Charging current >
5 A
Please wait for the battery to recharge before performing
the calibration step.
The battery indicator must be green to continue.
4:
No DC Current
Generators available.
No connected DC
Current Generators
are detected.
Ensure that all DC Current Generators are properly
connected.
5:
Mains Power missing.
The battery charger
is reporting a mains
power supply fault.
Ensure that the mains power is functioning normally.
6:
Charger error.
The battery charger
is reporting a general
error.
Ensure that the battery charger is functioning normally.
7:
Module temperature
too high.
The temperature of a
DC Current Generator
is too high.
Please wait for the DC Current Generator to cool before
proceeding.
8:
Ultracapacitor
temperature too high.
The temperature of
an ultracapacitor is
too high.
Please wait for the ultracapacitor to cool before proceeding.
10:
Battery voltage low
(warning).
Battery voltage < 14.5
VDC
Please wait for the battery to recharge before performing
the calibration step. The battery indicator must be green to
continue.
12.3.2 Initialisation warnings.
Warning no.
Cause:
Action:
1:
Battery voltage
critically low.
Battery voltage < 13
VDC
Please wait for the battery to charge fully before starting the
calibration. The battery indicator must be green to continue.
www.megger.com
137
Calibration