This means the input factor was unable to be determined within the valid range, or a time-out occurred.
Please check the following:
The calibration sensor is operational and properly connected to the µΩ input.
If problems persist, please contact Megger technical assistance.
12.2.6 Step 4/6 - LINEARITY ADJUSTMENT I-OUT
This dialog adjusts the linear calibration performed earlier for improved output accuracy.
The adjustments are calibrated using 6 injections at pre-defined points across the current range.
Press the ‘Start’ button to perform the calibration for a given point. When finished successfully, the button will be
highlighted in green and the resulting adjustment is displayed.
Repeat the procedure until all points are calibrated.
Press ‘Next’ to continue.
The linearity adjustment can be skipped to perform a faster calibration. However, to obtain maximum accuracy, it is
recommended to perform all 6 steps.