Message
Description
Appears in
Route has not been started
User tried to close a route without
testing any assets.
Executing routes.
Route close failed
User tried to close a route without
testing any assets.
Executing routes.
No asset selected
User tried to run an automatic or
sequenced test without selecting an
asset.
All automatic mode and testing assets
in sequenced mode.
No asset selected or selected asset has
no test configuration assigned
Testing in Automatic mode requires
an asset selection and an active test
configuration assigned.
All automatic mode tests.
No asset is selected or selected asset
has no sequence or test configuration
assigned
Testing in Sequenced mode requires
an asset selection, an active test
configuration, and an assigned
sequence.
Sequenced mode.
Asset being tested has no sequence
assignment
Testing in Sequenced mode requires
an asset selection, an active test
configuration, and an assigned
sequence.
Sequenced mode.
Invalid connection information
User tried to change Realm connection
using incorrect information.
Realm connection screen.
Connection Failed
User tried to change Realm connection
using incorrect information.
Realm connection screen.
Asset Exists in Another Installation.
Move Asset?
User tried to assign an asset to an
installation, but the software detected
that the asset is assigned to another
installation.
Installation editor.
Test results could not be saved
User tries to save test results during
surge and coil testing. Indicates a
potential out-of-memory issue.
Surge and coil tests.
Test Results could not be found for this
asset
User tries to open results for an asset
that has not been tested.
Route Execute screen.
Pulse-to-Pulse EAR+ limit has been
exceeded
PPEAR+ test results exceeded limit
defined in Test Config.
All surge tests.
Line-to-Line EAR+ limit has been
exceeded
L-L EAR+ test results exceeded limit
defined in Test Config.
All surge tests.
Line-to-Reference EAR+ limit has been
exceeded.
L-R EAR+ test results exceeded limit
defined in Test Config.
All surge tests.
Warning Open Lead detected
The software detects that one or more
test leads presents measurements
indicating an open lead condition.
All RLC tests.
Unable to hit target voltage on current
load. ADX energy is maxed out.
User tried to surge test an asset that
requires more energy than the tester
can provide.
All Surge tests.
Installation contains at least one asset.
Remove all assets before deleting.
User tried to delete an installation that
has one or more assets assigned to it.
Installation editor.
Installation is being used in route.
User tried to delete an installation that
is assigned to one or more routes.
Installation editor.
Active routes exist for _________
User tried to delete an asset that has
been assigned to one or more routes.
Asset editor.
Asset is being used in route
User tried to delete an asset that has
been assigned to one or more routes.
Asset editor.
www.megger.com
Megger Baker ADX User Guide
144
Error and Failure Messages List
Summary of Contents for ADX15
Page 1: ...Megger Baker ADX Automated Diagnostic Test Equipment User Guide...
Page 2: ...www megger com Megger Baker ADX User Guide...
Page 10: ...www megger com Megger Baker ADX User Guide viii Table of Contents...
Page 44: ...www megger com Megger Baker ADX User Guide 34...
Page 84: ...www megger com Megger Baker ADX User Guide 74 Routes...
Page 148: ...www megger com Megger Baker ADX User Guide 138...
Page 172: ...www megger com Megger Baker ADX User Guide 162...