7. The RLC tests will run first in this example sequence. The test runs quickly and you will see very little as the test
executes. If capacitance testing is enabled in the test configuration, the software will alert you to reconfigure the
test leads as shown in the example below.
8. Test Lead 1 will be connected to motor lead 1, test lead 2 to the motor frame, and test lead three will be
disconnected (not contacting any motor lead or ground).
NOTE: Acknowledge this message only after the leads have been reconfigured.
Pressing the OK button will immediately initiate the capacitance test.
Fig 122:
Auto RLC Test; leads reconfiguring message for capacitance testing.
9. After the capacitance test completes, another message appears letting you know that the
capacitance test has completed and that you must restore the test leads to their original connections
before continuing on with the next test in the sequence
.
CAUTION: Acknowledge this message only after the leads have been reconfigured. After pressing the
Continue Testing button, the software will immediately continue with the next test in the sequence—
often DC Tests.
Fig 123:
Auto RLC Test; leads reconfiguring message for capacitance testing.
www.megger.com
Megger Baker ADX User Guide
125
Sequenced Tests Mode
Summary of Contents for ADX15
Page 1: ...Megger Baker ADX Automated Diagnostic Test Equipment User Guide...
Page 2: ...www megger com Megger Baker ADX User Guide...
Page 10: ...www megger com Megger Baker ADX User Guide viii Table of Contents...
Page 44: ...www megger com Megger Baker ADX User Guide 34...
Page 84: ...www megger com Megger Baker ADX User Guide 74 Routes...
Page 148: ...www megger com Megger Baker ADX User Guide 138...
Page 172: ...www megger com Megger Baker ADX User Guide 162...