Table 17:
Instrument rating summary.
Parameter
Variant / Option
Value
Internal and operating environment
Pollution degree 2
Operating altitude
≤ 3,000 m (9,842 ft)
Operating temperature
5–40 °C (41–104° F)
Operating humidity
≤ 80% RH for temperature up to 31°C (88° F),
decreasing linearly to 50% RH at 40° C (104° F).
Storage temperature
0–60° C (32–140° F)
Ensure that the unit has sufficient time to warm to
ambient temperature before operating after storing
the unit in a colder area.
Storage humidity
Less than 95% non-condensing.
IP Rating
IP40
Mains Power Input
90–264 VAC, 47–63 Hz, 2.5 A, CAT II 300 V
Measurement connection rating
16 kV DC (nominal 15 kV)
Maximum generated voltage
Peak voltage for AC or DC test
ADX4
Nominal 100 V–4 kV
ADX6
Nominal 100 V–6 kV
ADX12
Nominal 100 V–12 kV
ADX15
Nominal 100 V–15 kV
ADX15A
Nominal 100 V–15 kV
Nominal 10 V–2kV armature output
Maximum input voltage rating
Must be connected only to isolated, de-energized
circuits.
See Caution below.
Standard Kelvin 4-wire test leads voltage rating
16 kV DC peak
CAUTION: The ADX must be connected only to isolated, de-energized circuits. Connection to live
circuitry can expose personnel to severe electrical shock risk, permanently damage tester, and void
warranty. Refer to chapter 1, “General Operating and Safety Information” for complete information on
safely connecting and operating the unit.
www.megger.com
Megger Baker ADX User Guide
166
General Specifications
Summary of Contents for ADX15
Page 1: ...Megger Baker ADX Automated Diagnostic Test Equipment User Guide...
Page 2: ...www megger com Megger Baker ADX User Guide...
Page 10: ...www megger com Megger Baker ADX User Guide viii Table of Contents...
Page 44: ...www megger com Megger Baker ADX User Guide 34...
Page 84: ...www megger com Megger Baker ADX User Guide 74 Routes...
Page 148: ...www megger com Megger Baker ADX User Guide 138...
Page 172: ...www megger com Megger Baker ADX User Guide 162...