Test Configuration
Surge Tests Section
In the Surge Tests section, Surge and PD tests can be enabled, and values can be provided for test voltages, EAR limits,
and more. Line-to-line/Reference EAR+™ limits can be applied to individual leads (phases) or to all leads. Pulse-to-pulse
EAR+™ limits and coil reference limits are always applied to all leads.
Fig 47:
Test Configuration screen; RLC section.
The ADX allows you to select up to three reference waveforms for your surge tests. Due to potential differences
between phases, each reference waveform should be assigned to a specific phase (A, B, or C) of a known good Asset
to ensure that each waveform for the device under tests is compared to a valid reference. If you use only one surge
reference, it will be applied to all surge waveforms collected.
If you create a test configuration for single coil testing, two references can be used—one for each direction.