Power Cord Wire Color Code:
Color
Definition
220V Cord
Brown
LINE
Blue
NEUTRAL
Green/Yellow
SAFETY
110V Cord
Black
LINE
White
NEUTRAL
Green
SAFETY
Instrument-Computer Ground Potential Check:
Perform this check if the power cord wiring is being changed, such as for replacement of the cord or if
changing the plug.
The instrument serial port ground connection is referenced to the instrument’s internal frame and safety
ground. Before connecting the communication cable to the instrument serial port, apply power to the
instrument and the host computer. Check the ground potential (voltage) between the serial port ground
pins on the computer and the instrument. The ground connection is pin 5 on DB-9 connectors and pin 7
on DB-25 connectors. Voltages in excess of 5VAC at 110V input power, or in excess of 10VAC at
220V input power may indicate a difference in ground wiring and can damage the instrument and/or the
computer. Check the wiring and take other steps as needed to reduce this difference before connecting
the communications cable. You may also use a data optoisolator.
Conditions of Illumination and Viewing
Geometry
Diffuse illumination, 8° viewing using a 6” (152.44 mm) plastic
integrating sphere coated with Spectraflect
Illumination
Xenon lamp with 1,000,000 flashes minimum life
UV-absorbing filter for 400-nm cutoff and control
Monochromator
Diffraction grating, dual channel
Detector
Dual diode array detectors
Interface Mode
Serial RS-232, USB
Area of View
Large area view port size: 1” (25.4 mm) diameter
Large area view measured area: 3/4” (19.0 mm) diameter
Small area view port size: 3/8” (9.5 mm) diameter
Small area view measured area: 1/4” (6 mm) diameter
Measurement Modes
Reflectance - Specular Included (RSIN)
Reflectance - Specular Excluded (RSEX)
Summary of Contents for EasyMatch QC
Page 11: ...Part I Software Overview...
Page 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Page 112: ...Part II Software Menu Commands...
Page 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Page 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Page 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Page 227: ...Part III Instruments...
Page 290: ...Part IV Reference...
Page 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...