EasyMatch QC User’s Manual Version 2.1
The Sensor Menu
8-19
As instructed, make sure that the standard port plate for your instrument is currently installed. This will
generally be the port insert with the largest opening that is not covered by glass. Click
OK
when this
port plate is installed.
Follow the on-screen prompts to perform a normal standardization in RTRAN mode. When
standardization is complete, the following prompt will be shown.
Screw the didymium filter, in its threaded tune, into the matching threads at the lens. Shut the
transmission compartment door. Click
OK
when you have done so. The Didymium Filter Test screen
appears.
As instructed, enter the 430 nm and 570 nm transmission values from the Didymium Filter area of your
tile data sheet. For an UltraScan PRO, also enter the 820 nm reading. Click
Accept
to enter these
values. (Click
Accept
on future instances of the test to acknowledge the correct values). Twenty
readings of the didymium filter will be made and then averaged.