EasyMatch QC User’s Manual Version 2.1
11G-2
UltraScan VIS
UltraScan VIS Features
The UltraScan VIS possesses the following default features:
Reflectance Port
The reflectance port is located at the front of the sensor. The port is covered by a spring-loaded sample
clamp which is used to hold the specimen in place during measurement. Two guidance holes are also
present to aid in placement of the white tile and light trap at the reflectance port. When the small area
view lens is in use, the small area port plate must be installed at the reflectance port.
Small Area View
The small area view feature includes a small area view lens and port plate. To use the small area of
view:
1. Remove the original port plate from the reflectance port and replace it with the small area port plate.
2. Standardize the instrument. Be sure to select 0.375 inch (9.525 mm) as the area view in the Setup
Mode window in EasyMatch QC. The lens will automatically adjust.
3. Place the sample at the reflectance port.
4. Be sure that the sample completely covers the reflectance port. (Use the retroviewer to confirm.)
Adjust the position of the sample if necessary.
The reflectance
port with the large
area view port
plate installed
Summary of Contents for EasyMatch QC
Page 11: ...Part I Software Overview...
Page 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Page 112: ...Part II Software Menu Commands...
Page 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Page 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Page 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Page 227: ...Part III Instruments...
Page 290: ...Part IV Reference...
Page 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...