EasyMatch QC User’s Manual Version 2.1
Measurement Values
A-17
this chapter. Its range is approximately 1 to 11 units, with the more yellow, pale worts at the low end of
the scale and the redder color of dark worts, beers, and caramels at the upper end of the scale.
Applicable equations are given below.
(
)
nm
430
10
T
log
-
*
1.27
*
10
ASBC
=
where
T
430 nm
= measured transmittance value (%) at 430 nm
and
Turbidity = free (“0.0”) if
(
)
nm
430
10
nm
700
10
T
log
-
*
1.27
*
0.039
100
T
log
-
*
1.27
≤
, else turbid (“1.0”).
where
T
700 nm
= measured transmittance value (%) at 700 nm.
ASBC Difference is also available, and is defined as follows:
dASBC-10 mm = (ASBC-10 mm)
sample
- (ASBC-10 mm)
standard.
ASTM D1500-33 mm (ColorQuest II Sphere, ColorQuest XE, ColorQuest XT,
UltraScan PRO, UltraScan XE, and UltraScan VIS only)
The ASTM D1500 scale is calculated as described below.
D1500 = 0.25 + 0.8695 (DX + DY + DZ)
where
=
98.078
X
log
-
DX
=
100.00
Y
log
-
DY
if Z > 0.01,
=
118.24
Z
log
-
DZ
if Z
≤
0.01,
=
118.24
0.01
log
-
DZ
X is the X tristimulus value for the sample calculated for C/2° conditions
Y is the Y tristimulus value for the sample calculated for C/2° conditions
Summary of Contents for EasyMatch QC
Page 11: ...Part I Software Overview...
Page 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Page 112: ...Part II Software Menu Commands...
Page 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Page 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Page 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Page 227: ...Part III Instruments...
Page 290: ...Part IV Reference...
Page 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...