EasyMatch QC User’s Manual Version 2.1
11G-44
UltraScan VIS
7. Remove the screws and cover from the lamp housing using a Phillips-head screwdriver.
8. Unplug the white connector of the lamp assembly by squeezing the latches on the sides of the
connector.
9. Remove the lamp assembly and replace with a new assembly. Be sure not to touch the bulb of the
lamp, as fingerprints reduce lamp efficiency and weaken the glass. If a fingerprint is deposited,
remove it using a clean wipe and isopropyl alcohol.