EasyMatch QC User’s Manual Version 2.1
11G-4
UltraScan VIS
A magnetic white ceramic disk on the face of the clamp provides a consistent white background for
measurements. A non-slip black pad (also magnetic) is also included that can be used to replace the
white disk for applications that require a non-slip or dark backing. To change to the pad, simply pop out
the white disk and insert the non-slip pad.
Transmission Compartment
The transmission compartment located in the middle of the sensor is used for measuring the transmitted
color of transparent solids or liquids. The transmission compartment door should be closed while
standardizing and taking measurements
.
Specular Included/Excluded Port
The UltraScan VIS sphere contains a specular exclusion port door which is closed during specular
included and transmission measurements and opened during specular excluded measurements. The
position of the door is controlled through EasyMatch QC. See the UltraScan VIS Standardization
section of this chapter for more information on the various measurement modes.
1
Closing the transmission compartment door while making both reflectance and transmittance measurements is the best
practice for this instrument. For more information, see the note in the UltraScan VIS Standardization section.
Button to raise
and lower the
clamp arm
Button to move
the clamp forward
and back
Summary of Contents for EasyMatch QC
Page 11: ...Part I Software Overview...
Page 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Page 112: ...Part II Software Menu Commands...
Page 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Page 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Page 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Page 227: ...Part III Instruments...
Page 290: ...Part IV Reference...
Page 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...