EasyMatch QC User’s Manual Version 2.1
UltraScan VIS
11G-37
20. Configure the screen as shown above, then click
Next
. Return to Step 12 and proceed through Step
17.
21. Remove the tape covering the reflectance port and the exterior of the sphere in the transmission
compartment.
22. Place the desired port plate at the reflectance port and snap it into place.
23. Turn on the UltraScan VIS by switching the on/off switch on the back of the sensor to the on
position. Allow the instrument to warm up for two hours before standardizing and making
measurements.
Notice: Do not block the vents in the top cover of the UltraScan VIS or the instrument
may overheat.
On/off switch
Power cord
receptor
Jack for foot
switch
Serial
communications
port
USB
communications
port