303
Error Display
13.4 Error Display
If any of the following are displayed on the screen, check the corresponding reference page.
Error display
Description
Solution/Reference
REF VAL
Measurement value is outside the
guaranteed accuracy range.
Check the measuring range. (p. 274)
MEAS ERR
Measurement error.
Check if the measurement cable has been
disconnected or connected incorrectly.
If an error is still displayed, the instrument
may be damaged.
Contact your authorized Hioki distributor or
reseller.
DISP OUT
Measurement value is outside the
display range.
Check the display range.
Hi Z
The measurement result is higher
than the judgment reference set for
the Hi Z reject function.
Check the connection.
“Detecting OPEN During 2-terminal
Measurement (HIGH-Z Reject Function)”
LEV ERR
This is displayed when abnormal
detection level is detected while
detection level monitoring function
is enabled.
Check the connection.
“Monitoring the Detection Level (Detection
Level Monitoring Function)” (p. 175)
MEMORY FULL
This is displayed when the set
number of measurement results
have been stored in the instrument
memory.
Load measurement values stored in the
instrument memory with the memory function
or clear the memory.
“Saving Measurement Results (Memory
The internal temperature has
exceeded the operating range or
the cooling fan has stopped.
Switch OFF the power supply at once.
Check the installation environment of the
instrument.
Check the condition of the instrument’s
cooling fan. There is a possibility of failure.
Contact your authorized Hioki distributor or
reseller.
ERROR
The current consumption of the front
USB terminal exceeds 500 mA.
Use a different type of USB flash drive.
?
Format of the USB flash drive is not
compatible with this instrument.
Use a different type of USB flash drive or
backup existing files in the USB flash drive
and format the drive before use.
File format error.
File cannot be loaded.
•
File is damaged.
•
File is not supported by this
instrument.
Check if file is corrupted or USB flash drive is
damaged.
Media space error.
USB flash drive does not sufficient
free memory space.
Use a different USB flash drive or increase
free space.
13
Maintenance and Service
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......