![Hioki IM7580 Instruction Manual Download Page 158](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306158.webp)
150
Calibration
Prevention of improper standard unit connection
To avoid improper connection of the three types of standard units (open/short/load), judgment can
be made by setting the limit with DC measurement.
During calibration measurement, if the limit values are set, exceeding the limit will result in an error,
and the calibration measurement will be stopped.
In case of an error, check that an appropriate standard unit is connected.
If the standard unit having a coaxial structure like the calibration kit is to be connected, connect
the standard unit by rotating the connector nut of the standard unit. The standard unit and the
central conductor of the connector will get damaged if the standard unit itself is rotated and
connected.
1
2
1
Press
[ADJUST]
.
2
Press
[Rdc LIMIT]
.
3
3
(1) Press
[OPEN MIN]
.
(2)
Set the limit values with the numeric
keypad*.
Settable range
-9.99999 G to 9.99999 G
(3)
Press
[SET]
to close the setting screen.
During open calibration measurement, an error occurs
if the DC measured value falls below this limit, and the
measurement is stopped.
4
4
(1) Press
[LOAD MAX]
.
(2)
Set the limit values with the numeric
keypad*.
Settable range
-9.99999 G to 9.99999 G
(3)
Press
[SET]
to close the setting screen.
During load calibration measurement, an error occurs
when the DC measured value exceeds this limit, and
the measurement is stopped.
Go to the next page.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......