![Hioki IM7580 Instruction Manual Download Page 133](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306133.webp)
125
Equivalent Circuit Analysis Function
4.9.2
Configuring Basic Settings for Analysis
(1) Setting the equivalent circuit model
Select the equivalent circuit model you wish to use for equivalent circuit analysis.
You will be able to estimate constants more accurately by selecting the appropriate equivalent
circuit model.
1
3
2
1
Press
[SETUP]
.
2
Press
[CIRCUIT]
tab.
3
Press
[MODEL]
.
4
5
6
4
Select the model to be used in equivalent circuit
analysis.
[OFF]
Turns off the equivalent circuit function.
[HOLD]
Selects the equivalent circuit model manually.
[AUTO]
Selects the most suitable equivalent circuit
model automatically.
5
(If
[HOLD]
is selected in step
4
)
Select an equivalent circuit model to be used.
6
Press
[EXIT]
to close the setting screen.
• When the equivalent circuit model A to E is selected, HOLD is automatically set.
•
Equivalent Circuit Model” (p. A5).
4
Analyzer Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......