Sweep Measurement
Setting
of sweep
parameters
(
[SOURCE]
)
Sweep range
setting
Contents of
setting
Settable range
Voltage
[V]
[START-STOP]
Start value of
sweep
[START]
4 mV to 1001 mV
End value of
sweep
[STOP]
4 mV to 1001 mV
Number of sweep
points
[NUM]
1 to 801
* Setting method for sweep points is fixed to
[LINEAR]
.
[INTVL MEAS]
Start value of
sweep
[POINT]
4 mV to 1001 mV
Measurement time
interval
[INTERVAL]
0.00000 s to 1000.00 s
Number of
measurements
[NUM]
1 to 801
*The measurement interval for INTERVAL measurement is
reflected in the point delay time.
Current
[I]
[START-STOP]
Start value of
sweep
[START]
0.09 mA to 20.02 mA
End value of
sweep
[STOP]
0.09 mA to 20.02 mA
Number of sweep
points
[NUM]
1 to 801
* Setting method for sweep points is fixed to
[LINEAR]
.
[INTVL MEAS]
Start value of
sweep
[POINT]
0.09 mA to 20.02 mA
Measurement time
interval
[INTERVAL]
0.00000 s to 1000.00 s
Number of
measurements
[NUM]
1 to 801
4
Analyzer Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......