286
Measurement Accuracy
Accuracy of inductor Ls = 150 nH
Example: Measurement frequency = 100 MHz, measurement signal level=+1 dBm, measurement
speed = FAST
1
Z and
θ
of the sample are measured and we assume that the measurement values are as
follows.
Z = 94.292
Ω
θ = 88.25°
2
Obtain Ea.
From measurement conditions and accuracy specification:
Er = 0.09
Ea = 0.5 + Er = 0.59
3
Obtain Zs.
From measurement conditions and accuracy specification:
Zsr = 13.5
F = 100
(measurement frequency [MHz])
With the above, Zs is calculated as follows:
Zs = 20 + Zsr + 0.5
×
F
= 20 + 13.5 + 0.5
×
100
= 83.5
[m
Ω
]
4
Obtain Yo.
From measurement conditions and accuracy specification:
Yor = 7.5
F = 100
(measurement frequency [MHz])
With the above, Yo is calculated as follows:
Yo = 30 + Yor + 0.15
×
F
= 30 + 7.5 + 0.15
×
100
= 52.5
[µS]
5
Obtain Eb with Zs, Yo and measurement value Zx.
Eb
Zs
Zx
Yo Zx
=
+
⋅
×
100
=
×
+
×
×
83 5
1000
1
94 292
52 5
1000000
94 292 100
.
.
.
.
[%]
= (0. 0.004950) × 100
= 0.5836
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......