![Hioki IM7580 Instruction Manual Download Page 291](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306291.webp)
283
Measurement Accuracy
12.2 Measurement Accuracy
12.2.1 Impedance Measurement
Accuracy guaranteed
temperature and
humidity range
0°C to 40°C (32°F to 104°F), 20% RH to 80% RH (no condensation)
* 30°C (86°F) or more, wet-bulb temperature 27°C (80.6°F) or less
However, within ±5°C (±9°F) of the calibration temperature.
Calibration temperature
range
Based on the operating temperature of calibration kit.
Guaranteed accuracy
period
1 year (Ensure open/short/load calibration are performed daily before
measurement.)
Warm-up time
At least 60 minutes
Calibration face
7-mm terminal face of Model IM9906 Adapter (3.5 mm/7 mm) attached to the 3.5-
mm terminal of test head
(After performing open/short/load calibration with calibration kit)
Measurement
conditions
Frequency, power and speed points where open/short/load calibration was
implemented
Calibration kit
When products with the following specifications or equivalent are used
LOAD (50
Ω
): VSWR=1.005 max
OPEN: Reflection coefficient 0.995 max
SHORT: Reflection coefficient 0.995 max
Measurement accuracy
Z: ±(Ea + Eb) [%]
θ
: ±0.58 × (Ea + Eb) [°]
Ea = 0.5 + Er
Er: Power = between -7 dBm and +7 dBm
Frequency
FAST
MED
SLOW
SLOW2
1 MHz to 100 MHz
0.09
0.06
0.036
0.03
100.01 MHz to 300 MHz
0.108
0.078
0.039
0.036
Er: Power = between -40 dBm and -7.1 dBm
Frequency
Er
Eoff
FAST
MED
SLOW
SLOW2
1 MHz to 100 MHz
3 × 10
(-0.046P + Eoff)
-1.8
-2
-2.15
-2.3
100.01 MHz to 300 MHz
3 × 10
(-0.048P + Eoff)
-1.75
-1.9
-2.1
-2.25
P: Power setting value [dBm]
Eb
Zs
Zx
Yo Zx
=
+
⋅
×
100
[%] (|Zx|: Measurement value of Z Unit: [
Ω
])
Zs
Zsr
F
=
+
+
×
(
)
20
0 5
1000
.
[
Ω
] (F: Measurement frequency [MHz])
12
Specifications
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......