Timing Chart
•
In the continuous measurement screen, comparator result output signals other than AND and panel
load signals (LD-VALID, LD0 to LD6) cannot be used.
Refer to “Continuous Measurement Function” (p. 161)
• Whether the comparator judgment results are reset when the signal changes to EOM (HIGH) or
updated when measurement is completed, can be selected on the instrument or by a communication
command. See
“8.6.4 Setting the EOM Output Method (EOM Mode)” (p. 218).
LCR Application Disk - Communication Commands (
:IO:RESult:RESET
).
•
For each time of other timing chart times, refer to “LCR mode” (p. 196).