General Specifications
Parameter
color change
function
This function enables display colors to be changed for measurement values.
Absolute
measurement
value display
function
Absolute measurement value display function for measurement values (
θ
and ∆% excluded)
Key-lock
function
Can be enabled and disabled by front panel key operation.
Key lock is released by entering a passcode.
Memory
function
32000 LCR measurements and 100 ANALYZER sweeps can be saved in this instrument.
(Reading via RS-232C, GP-IB, USB, LAN or USB flash drive is possible.)
Beep sound
Beep sound for the comparator judgment result (IN or NG) can be set to ON/OFF.
Beep sound for key input can be set to ON/OFF.
15 types of beep sounds are available.
I/O judgment
output
delay function
1.
Delay from judgment result output to EOM
0.00000 s to 0.99999 s
2.
Judgment result output reset timing modification function
IO trigger
1. This function enables trigger input during measurement.
2. Edge selection (rising, falling)
I/O EOM
EOM signal output method (pulse, hold)
0.00001 s to 0.99999 s
Warm-up
function
A message appears 60 min after power-on.
(6) Compensation
Open/short/load
calibrations
(compensation to
test head)
ALL and SPOT available, compensation value check, compensation value read/write
possible
(ANALYZER: SPOT only)
Number of SPOT compensations: 5 (LCR), 801 (ANALYZER)
Open/short
compensations
(compensation to
test fixture)
ALL and SPOT available, compensation value check, compensation value read/write
possible
Number of SPOT compensations: 5 (LCR), 801 (ANALYZER)
ALL compensation or SPOT compensation works with open/short/load calibration.
Electrical length
compensation
Compensation range: 0.00 mm to 100.00 mm
Correlation
compensation
Enter the compensation coefficients a and b for the following expression.
[Measurement value after compensation] = a × [measurement value] + b
Setting range for a: -999.999 to +999.999
Setting range for b: -9.99999 G to +9.99999 G
12
Specifications
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......