41
Setting Basic Settings of Measurement Conditions
3.2.8
Displaying Average Values (Average)
The measured values can be averaged using the averaging function. The variations in the
displayed measured values can be reduced with this function.
Internal trigger
The measured values are always the moving average from present
to before the number of times of averaging.
(When the sample to be tested is changed, it takes time for the
values to stabilize and the results are reliable.)
External trigger
It is the arithmetic mean of the number of times of averaging from
the trigger input.
When averaging is executed 4 times, the number of measurements, measurement output points,
and measured value calculation method during output are as follows.
Moving average
Measurement output values
Measurement points
Time
M4+M5+M6+M7
4
M5+M6+M7+M8
4
M6+M7+M8+M9
4
M7+M8+M9+M10
4
M7
M8
M9
M10
M1
1
M1+M2+M3
3
M1+M2+M3+M4
4
M2+M3+M4+M5
4
M3+M4+M5+M6
4
M1
M2
M3
M4
M5
M6
M1+M2
2
Arithmetic mean
M1
M2
M3
M4
M5
M6
M7
M8
M9
M10
Time
Measurement output values
Measurement points
M5+M6+M7+M8
4
M1+M2+M3+M4
4
3
LCR Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......