![Hioki IM7580 Instruction Manual Download Page 213](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306213.webp)
205
Timing Chart
8.2 Timing Chart
8.2.1
LCR Mode
If you set the judgment condition for the comparator (trigger setting is external trigger) and in that
state if a trigger signal is input from the EXT I/O or
[TRIG]
is pressed in the screen, the judgment
result is output from the signal line for comparator result output of the EXT I/O after measurement
is completed.
Furthermore, if the panel No. has been selected with the panel load signal when a trigger signal is
input from the EXT I/O, measurement is performed after the measurement condition of that panel
No. is loaded.
Examples of the measurement timing:
In the timing example, the valid edge of the TRIG
―
―
―
―
signal is set to falling (ON).
Chuck
OFF
OFF
ON
ON
ON
ON
Previous
judgment result
Judgment result
ON
OFF
t5
t6
t1
t2
t4
t3
t7
t8
Chuck
Release
TRIG
―
―
―
―
(Measurement start signal)
EOM
―――
(Measurement complete signal)
INDEX
―
―――
―
(Analog measurement complete)
ERR
―――
(Error output)
LD_VALID
―
―
―
―
―
―
―
―
LD0
―
―
―
to LD6
―
―
―
Connection status
(Judgment result output)
EOM:OFF From trigger to the end of measurement
INDEX:OFF Probe chuck period (Do not disengage the probe.)
Whether the judgment results of comparator and BIN judgment are reset when the signal
changes to EOM (HIGH) or updated when measurement is completed, can be selected on the
instrument or by a communication command.
See “8.6.3 Setting Reset of Judgment Results (Judgment Result Signal Reset)” (p. 217).
See LCR Application Disk - Communication Commands (
:IO:RESult:RESET
).
8
External Control
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......