Calibration
5.2.2
Setting the Electric Length [LENGTH]
Compensation is performed for the error caused by the phase shift occurring between the
calibration reference surface and the measurement sample connection surface.
Enter the electric length between the calibration reference surface on which open calibration, short
calibration, and load calibration were performed for the surface where the measurement sample is
connected.
The guaranteed accuracy range differs depending on the electric length.
See “Electrical length compensation” (p. 279).
1
1
Press
[ADJUST]
.
2
2
Press
[LENGTH]
.
The numeric keypad can be
used for input.
4
3
3
Set the electric length with
/
or the numeric
keypad.
(when the numeric keypad is used, press
[SET]
.)
[C]
Repeats the input.
[CANCEL]
Cancels the setting.
4
Press
[EXIT]
to close the setting screen.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......