149
Calibration
3
3
Press
[AC + DC]
.
Starts measurement.
If the standard unit used in AC measurement is
different from that the one used in DC measurement,
press
[AC]
,
[DC]
.
The results are displayed under
[SHORT]
after
completion of measurement.
[CANCEL]
Cancels the setting, and closes the screen.
Load measurement
2
1
Connect the standard unit for load to the test
sample connection terminal.
2
Press
[LOAD]
.
3
3
Press
[AC + DC]
.
Starts measurement.
If the standard unit used in AC measurement is
different from that the one used in DC measurement,
press
[AC]
,
[DC]
.
The results are displayed under
[LOAD]
after
completion of measurement.
[CANCEL]
Cancels the setting, and closes the screen.
5
Calibration and Compensation
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......