![Hioki IM7580 Instruction Manual Download Page 123](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306123.webp)
115
Judging Measurement Results (Comparator Function)
4.8.3
Setting the Judgment Area to Display in the Measurement
Screen
1
1
Press
[SETUP]
.
2
3
2
Press
[COMP]
tab.
3
Press
[AREA]
.
4
5
4
Select the parameter that will display the judgment
areas.
[PARA1]
Displays the judgment area for parameter 1.
[PARA2]
Displays the judgment area for parameter 2.
[PARA3]
Displays the judgment area for parameter 3.
[PARA4]
Displays the judgment area for parameter 4.
[OFF]
Judgment area is not displayed.
Selects the sweep parameter to display judgment area
on the second normal sweep graph if the graph display
setting is
[MULTI]
in the area setting of “GRAPH2”.
5
Press
[EXIT]
to close the judgment settings
screen.
6
Press
[EXIT]
to close the advanced settings
screen.
If it is difficult to see the judgment area of the graphical screen, increasing the brightness will
improve the visibility.
See “Setting the screen brightness” (p. 183)
4
Analyzer Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......