Timing Chart
•
If the trigger setting is set to STEP,
INDEX
―
―――
―
and EOM
―――
transition to LOW every time the measurement
for each point is completed, and transitions to HIGH if there is trigger input. ERR
―――
also transitions to
LOW each time measurement is completed if a measurement error occurs.
• Whether the judgment results of comparator measurement are reset at the time of the measurement
start signal or updated when measurement is completed, can be selected on the instrument or by a
communication command. See “8.6.4 Setting the EOM Output Method (EOM Mode)” (p. 218).LCR
Application Disk - Communication Commands (
:IO:RESult:RESET
).
•
For each time of other timing charts, refer to Refer to “8.2.1 LCR Mode” (p. 205).
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......