72
Setting Basic Settings of Measurement
4.2.4
Applying the Signal to the Sample during Measurement Only
(Trigger Synchronous Output)
This function enables the measurement signal to be output for only the initial sweep point after
measurement is triggered, so that the signal is applied to the sample during measurement only. You
can also set a delay time to ensure that data is acquired after the sample stabilizes.
Thus reducing the generation of heat in the sample and decreasing electrode wear.
Trigger synchronous output: OFF
Trigger synchronous output: ON
Measurement
Calculation
TRIG
―
―
―
―
EOM
―――
INDEX
―
―――
―
Signal
Measurement
Calculation
Synchronous
signal wait
INDEX delay
TRIG
―
―
―
―
EOM
―――
INDEX
―
―――
―
Signal
1
3
2
1
Press
[SETUP]
.
2
Press
[SWEEP]
tab.
3
Press
[SYNC]
.
4
5
4
Press
[SYNC]
.
5
Select ON/OFF for the trigger synchronous output.
[OFF]
Disables the trigger synchronous output.
[ON]
Enables the trigger synchronous output.
Go to the next page.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......