Common Functions (LCR Mode, ANALYZER Mode)
5
6
7
5
Press
[MEMORY]
.
6
Select
[ON]
/
[IN]
/
[OFF]
from the memory function.
If the comparator and BIN functions are not set, the
operation of IN and ON will be the same.
[OFF]
Disables the memory function.
[IN]
(LCR only)
Saves the measured values to the memory
only when a pass judgment is made for
all of the parameters judged with the
comparator and BIN functions.
(The measured values are not saved even
if one of the comparator results is
HI
,
LO
,
or the BIN result is
OUT-OF-BINS
.)
[ON]
Saves all measured values to memory.
[CLEAR]
Clears all the measured values saved in
the instrument memory.
[SAVE]
Saves the measured values stored in the
instrument memory to a USB flash drive
and then clears the measured values from
the instrument memory.
The measured values are saved to the
“MEMORY” folder in the USB flash drive.
The file name is automatically assigned
from the date and time.
7
Press
[EXIT]
to close the setting screen.
•
If the memory function is enabled (
[ON]
/
[IN]
), the number of memory items currently saved is
displayed in the measurement screen.
See
“13.4 Error Display” (p. 303).
• Save the measurement results stored in the instrument to a USB flash drive or acquire them with the
:MEMory?
command.
•
The internal data is lost if the memory function setting is changed.
•
The following message appears on the measurement screen when the instrument memory becomes
full. If this message appears, subsequent measured values will not be saved. To resume saving,
transfer or clear the measured values from the instrument memory.
See “13.4 Error Display” (p. 303).
•
If the contact check function has been set, the measured value will not be saved if all four of the
following conditions apply.
Refer to “7.1 Checking Contact Defects and the Contact State (Contact Check Function)” (p. 169)
- If the contact check timing has been set to
[BEFORE]
or
[BOTH]
.
- If LIMIT has been set.
- If ERR ABORT has been set to ON.
- If LIMIT judgment detected an error at the timing of BEFORE.
7
Application Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......