Error Compensation
5.3 Error Compensation
5.3.1
Setting the Compensation [COMPEN]
The errors between the calibrated calibration reference surface and the measurement terminal are
eliminated.
When the test sample is connected to the measurement terminal extended from the calibration
reference surface on which open calibration, short calibration, and load calibration were performed,
perform the measurement when the terminal to connect the test sample is shorted and opened
respectively.
1
1
Press
[ADJUST]
.
2
2
Press
[COMPEN]
.
4
3
3
Press
[MODE]
.
4
Selects the compensation method.
[OFF]
Compensation is not performed.
[ON]
Performs compensation. The points of
compensation are the same as those of
[CAL]
calibration.
5
6
5
Select the input parameter pattern for the
reference value set by
[OPEN]
and
[SHORT]
respectively.
See “3.2.1 Setting Display Parameters” (p. 32).
[OPEN]
G-Cp, G-B
[SHORT]
Rs-Ls, Rs-X
6
Press
[EXIT]
to close the setting screen.
5
Calibration and Compensation
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......