113
Judging Measurement Results (Comparator Function)
4
5
6
4
Select the judgment mode.
[OFF]
Disables comparator function.
[AREA]
Enables area judgment. (p. 116)
[PEAK]
Enables peak judgment. (p. 120)
5
When a measured value is outside the guaranteed
accuracy range, set how the measured value is
judged.
[DO]
Judges when a measured value is outside
the guaranteed accuracy range.
[NOT]
Outputs error for HI judgment when a
measured value is outside the guaranteed
accuracy range.
6
Sets the beep sounds for judgment results.
[OFF]
No beep sound.
[IN]
Beeps if all the judgment results are IN.
[NG]
Beeps if one of the judgment results is LO
or HI.
7
8
9
7
With
/
, set the beep tone.
Settable range 0 to 14
8
With
/
, set the beep volume.
Settable range 1 to 3
9
Press
[EXIT]
to close the judgment settings
screen.
10
Press
[EXIT]
to close the advanced settings
screen.
4.8.2
Setting the Parameter to be Judged
1
3
2
1
Press
[SETUP]
.
2
Press
[COMP]
tab.
3
Press
[PARA]
.
4
Analyzer Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......