244
Saving Data to USB Flash Drive
Save examples:
Settings: DATE: ON, SET: ON, PARA: ON, DELIM: “ , “ (comma), QUOTE: “ (double quotation mark)
“HIOKI E.E. CORPORATION”,”IM7580”,”Ver. 1.00”
“Serial No. 123456789”
“DATE”,”14-05-08”
“TIME”,”15:17:10”
“TRIG”,”EXT”
“TRIG DELAY”,”0.00000”,”s”
“SYNC”,”OFF”,””
“TRIG SYNC WAIT”,”0.00100”,”s”
“TRIG SYNC INDEX WAIT”,”0.00000”,”s”
“FREQ”,” 1.0000E+06”,”Hz”
“POWER”,”0.0”,”dBm”
“SPEED”,”MED”,””
“AVG”,”001”,””
“RDC”,”OFF”
“WAVE”,”0001”
“DC WAIT”,”0.00000”,”s”
“AC WAIT”,”0.00000”,”s”
“SCALING”,”OFF”
“Z[ohm]”,”OFF”,”PHASE[deg]”,”OFF”
“5.00”,””,”175.604”,””
In case of LCR mode
“HIOKI E.E. CORPORATION”,”IM7580”,”Ver. 1.00”
“Serial No. 123456789”
“DATE”,”14-05-08”
“TIME”,”15:17:16”
“SOURCE”,”FREQ”
“TRIG”,”SEQ”
“TRIG DELAY”,”0.00000”,”s”
“TRIG SYNC”,”OFF”
“TRIG SYNC WAIT”,”0.00100”,”s”
“TRIG SYNC INDEX WAIT”,”0.00000”,”s”
“RDC”,”OFF”
“WAVE”,”0001”
“DC WAIT”,”0.00000”,”s”
“AC WAIT”,”0.00000”,”s”
“SCALING”,”OFF”
“No.”,”FREQ[Hz]”,”LEVEL”,””,”SPEED”,”AVG”,”DELAY”,”Z[ohm]”,”PHASE[deg]”,”Rs[ohm]”,”X[ohm]”
“1”,” 1.0000E+06”,”0.0”,”dBm”,”MED”,”001”,”0.00000s”,”5.00”,”175.598”,”-5.00”,”459.52E-03”
“2”,” 1.0289E+06”,”0.0”,”dBm”,”MED”,”001”,”0.00000s”,”6.00”,”175.729”,”-5.00”,”447.03E-03”
“3”,” 1.0587E+06”,”0.0”,”dBm”,”MED”,”001”,”0.00000s”,”6.00”,”175.858”,”-6.00”,”434.69E-03”
“4”,” 1.0893E+06”,”0.0”,”dBm”,”MED”,”001”,”0.00000s”,”6.00”,”175.982”,”-6.00”,”422.57E-03”
“5”,” 1.1208E+06”,”0.0”,”dBm”,”MED”,”001”,”0.00000s”,”6.00”,”176.100”,”-6.00”,”411.20E-03”
“6”,” 1.1533E+06”,”0.0”,”dBm”,”MED”,”001”,”0.00000s”,”6.00”,”176.217”,”-6.00”,”399.83E-03”
“7”,” 1.1866E+06”,”0.0”,”dBm”,”MED”,”001”,”0.00000s”,”6.00”,”176.324”,”-6.00”,”389.28E-03”
In case of ANALYZER mode
“HIOKI E.E. CORPORATION”,”IM7580”,”Ver. 1.00”
“Serial No. 123456789”
“DATE”,”14-05-08”
“TIME”,”15:21:57”
“LCR”,”1”,”1405081406”
“TRIG”,”EXT”
“TRIG DELAY”,”0.00000”,”s”
“SYNC”,”OFF”,””
“TRIG SYNC WAIT”,”0.00100”,”s”
“TRIG SYNC INDEX WAIT”,”0.00000”,”s”
“FREQ”,” 1.0000E+06”,”Hz”
“POWER”,”0.0”,”dBm”
“SPEED”,”MED”,””
“AVG”,”001”,””
“RDC”,”OFF”
“WAVE”,”0001”
“DC WAIT”,”0.00000”,”s”
“AC WAIT”,”0.00000”,”s”
“SCALING”,”OFF”
“Z[ohm]”,”OFF”,”PHASE[deg]”,”OFF”
“5.00”,””,”175.605”,””
In case of CONTINUOUS measurement mode
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......