![Hioki IM7580 Instruction Manual Download Page 188](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306188.webp)
180
Common Functions (LCR Mode, ANALYZER Mode)
7.3 Common Functions (LCR Mode, ANALYZER
Mode)
These settings are common for LCR mode and ANALYZER mode.
These settings provide the same conditions to both modes.
7.3.1
Saving Measurement Results (Memory Function)
The measurement results can be saved in the instrument (Up to 32,000 items for LCR, and 100 sweeps
for ANALYZER).
The saved measurement results can be saved to a USB flash drive.
See “11.4.4 Saving Memory Data” (p. 258).
Saved data can also be acquired using a communication command.
The items saved to the memory are in accordance with the
:MEASure:VALid
setting. For details on
how to acquire the saved measurement results or how to set
:MEASure:VALid
, refer to LCR Application
Disk (Communication Commands).
1
2
3
1
Press
[SETUP]
.
2
Press
[COMMON]
tab.
Press
[ADVANCED]
tab for the ANALYZER mode.
3
Press
[MEMORY]
.
4
The numeric keypad can be used for input.
4
Set the number of measurement results to save
with
/
.
Settable range 1 to 32,000 (LCR mode)
100 fixed (ANALYZER mode)
Go to the next page.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......