![Hioki IM7580 Instruction Manual Download Page 183](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306183.webp)
175
Checking Contact Defects and the Contact State (Contact Check Function)
7.1.4
Monitoring the Detection Level (Detection Level Monitoring
Function)
This function can detect the abnormal measurement waveforms when there is contact between the
test sample and the instrument by monitoring the variations in the r.m.s. value of voltage and the
r.m.s. value of current. During analog measurement, the r.m.s. value of voltage and the r.m.s. value
of current are calculated several times.
The first-calculated r.m.s. value of voltage and the r.m.s value of current are considered as the
reference value respectively. The percentage
∆
% of subsequent r.m.s. values of voltage and
current relative to the reference value is computed using the following formula.
∆
% =
(r.m.s. value - Reference value)
Reference value
× 100 [%]
An error is detected when
∆
% is higher than the set limit value.
1
2
3
1
Press
[SETUP]
.
2
Press
[CONTACT]
tab.
Press
[SWEEP]
tab for the ANALYZER mode.
3
Press
[LEV CHECK]
.
4
5
4
Press
[CHECK]
.
5
Select ON/OFF for the detection level monitoring
function.
[OFF]
Sets the detection level monitoring function to
OFF.
[ON]
Sets the detection level monitoring function to
ON.
7
Application Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......