General Operation
Chapter 1: Introduction
Advanta™ Bed Service Manual (man168ra)
Page 1 - 33
1
For B model beds:
•
Ensure that the patient is centered on the mattress and lying still and
that no visitors or caregivers are inadvertently moving the bed or
affecting the bed scale, such as resting a hand or arm on a siderail or
headboard.
•
Press and hold the Position switch until the indicator light next to the
word Position starts to flash.
•
Release the Position switch.
It takes approximately 5 seconds for the PPM to activate.
If patient movement is detected while in the Position mode, an audible alarm
sounds from the bed, the top Position light begins flashing, and a repeating
nurse call is sent. If the patient leaves the bed, the Out of Bed light begins
flashing. Both lights and alarms continue until the PPM system is turned off,
even if the patient lies down on the bed again.
Exiting mode: The Exiting mode sets off an alarm when a patient moves both
away from the head section and towards the egress points of the bed. This
mode should be used when a caregiver wants to be alerted when egress is
attempted.
To activate:
For A model beds:
•
Ensure that the patient is centered on the mattress and lying still, and
that no visitors or caregivers are inadvertently moving the bed or
affecting the bed scale such as resting a hand or arm on a siderail or
headboard.
•
Press and hold the enable switch, and press and release the mode
switch until the indicator light next to the word Exiting illuminates.
•
Release the enable switch.
It takes approximately 5 seconds for the PPM to activate.
For B model beds:
•
Ensure that the patient is centered on the mattress and lying still, and
that no visitors or caregivers are inadvertently moving the bed or
Summary of Contents for Advanta P1600
Page 2: ......
Page 6: ...Revisions Page iv Advanta Bed Service Manual man168ra NOTES...
Page 22: ...Chapter 1 Introduction Page 1 4 Advanta Bed Service Manual man168ra NOTES...
Page 142: ...Chapter 3 Theory of Operation Page 3 4 Advanta Bed Service Manual man168ra NOTES...
Page 310: ...Warranty Chapter 5 Parts List Page 5 4 Advanta Bed Service Manual man168ra NOTES...
Page 353: ...Footboard P1606 Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 47 5 NOTES...
Page 371: ...Power Module Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 65 5 NOTES...
Page 386: ...Chapter 6 General Procedures Page 6 2 Advanta Bed Service Manual man168ra NOTES...
Page 404: ...Chapter 7 Accessories Page 7 2 Advanta Bed Service Manual man168ra NOTES...
Page 412: ...Schematic Wiring Diagram Patient Switch Board P N 67208 FO 3 2 m168a103 Back to Chapter 3...
Page 413: ...Schematic Wiring Diagram J Box UTV Board P N 64935 FO 3 3 m168_107 Back to Chapter 3...
Page 417: ...Schematic Wiring Diagram Logic Board PN 63285 FO 3 7 m168_115 Back to Chapter 3...
Page 418: ...Schematic Wiring Diagram Power Board PN 63349 FO 3 8 m168_118 Back to Chapter 3...
Page 419: ...Schematic Wiring Diagram Frame Interface Board PN 63352 FO 3 9 m168_120 Back to Chapter 3...
Page 422: ...Schematic Wiring Diagram Sensor Board PN 4474001 FO 3 12 m168_128 Back to Chapter 3...
Page 425: ...Schematic Wiring Diagram Head Sensor Assembly P N 63888 FO 3 14 m168a186 Back to Chapter 3...
Page 427: ...Schematic Wiring Diagram HIlow Knee Sensor PN 63885 FO 3 16 m168_178 Back to Chapter 3...