Introduction
Chapter 1: Introduction
Page 1 - 14
Advanta™ Bed Service Manual (man168ra)
raising the siderail, you will hear a “click” when the siderail is securely in
place. To store a siderail, lower it completely, then push it in under the bed
surface.
Always tug slightly on a siderail after adjusting it to ensure that it is securely
locked in place.
ZoneAire® Sleep Surface
The ZoneAire® Sleep Surface is a multiple-zoned mattress enclosed in foam,
ideally suited for the reduction of patient interface pressure. The closed air
system is monitored electronically to ensure proper inflation. The system
operates in three modes: pressure relief, auto firm, and heel relief.
Pressure Relief Mode
The pressure relief mode is for patients who are assessed to be at risk for
pressure ulceration, and who require pressure relief. In the pressure relief
mode, the air pressure in each zone of the mattress is automatically maintained
to optimize the interface pressure on the patient. Air volumes change to control
the surface around the patient’s body, especially the bony areas, to evenly
distribute the pressure across the body. The surface adjusts to the patient’s
weight distribution, position, and the bed articulation.
To activate: Press the surface power switch to begin the automatic pressure
relief mode.
Auto Firm Mode
The auto firm mode inflates the air mattresses to maximum firmness. This
helps with patient transfers, procedures, and repositioning. The auto firm mode
also comes on automatically when the CPR Release is activated. After 30
minutes, the mattress returns to its original setting. This helps ensure that a
patient does not rest on a firm surface for an extended period of time.
To activate: Press and release the auto firm switch on the ZoneAire® Sleep
Surface controls. The indicator light illuminates, showing that the mattress is
in the auto firm mode.
Heel Relief Mode
The heel relief mode is for patients who are assessed to be at risk for pressure
ulceration, with particular risk to the heel area due to their low level of
mobility. The heel relief mode decreases the pressure under the heel area, and
slightly raises the surface under the calves. There are three heel relief zones
Summary of Contents for Advanta P1600
Page 2: ......
Page 6: ...Revisions Page iv Advanta Bed Service Manual man168ra NOTES...
Page 22: ...Chapter 1 Introduction Page 1 4 Advanta Bed Service Manual man168ra NOTES...
Page 142: ...Chapter 3 Theory of Operation Page 3 4 Advanta Bed Service Manual man168ra NOTES...
Page 310: ...Warranty Chapter 5 Parts List Page 5 4 Advanta Bed Service Manual man168ra NOTES...
Page 353: ...Footboard P1606 Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 47 5 NOTES...
Page 371: ...Power Module Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 65 5 NOTES...
Page 386: ...Chapter 6 General Procedures Page 6 2 Advanta Bed Service Manual man168ra NOTES...
Page 404: ...Chapter 7 Accessories Page 7 2 Advanta Bed Service Manual man168ra NOTES...
Page 412: ...Schematic Wiring Diagram Patient Switch Board P N 67208 FO 3 2 m168a103 Back to Chapter 3...
Page 413: ...Schematic Wiring Diagram J Box UTV Board P N 64935 FO 3 3 m168_107 Back to Chapter 3...
Page 417: ...Schematic Wiring Diagram Logic Board PN 63285 FO 3 7 m168_115 Back to Chapter 3...
Page 418: ...Schematic Wiring Diagram Power Board PN 63349 FO 3 8 m168_118 Back to Chapter 3...
Page 419: ...Schematic Wiring Diagram Frame Interface Board PN 63352 FO 3 9 m168_120 Back to Chapter 3...
Page 422: ...Schematic Wiring Diagram Sensor Board PN 4474001 FO 3 12 m168_128 Back to Chapter 3...
Page 425: ...Schematic Wiring Diagram Head Sensor Assembly P N 63888 FO 3 14 m168a186 Back to Chapter 3...
Page 427: ...Schematic Wiring Diagram HIlow Knee Sensor PN 63885 FO 3 16 m168_178 Back to Chapter 3...