Preventive Maintenance
Chapter 6: General Procedures
Advanta™ Bed Service Manual (man168ra)
Page 6 - 11
6
Siderail frame
Test the siderail for proper latching.
Pivot points
Lubricate all pivot points on the bed.
CPR release
Test the CPR release for proper operation and reset of the head
screw drive.
Adjust the CPR release if necessary (see “CPR Release Handle”
on page 4-79).
Night light
Check the night light to ensure that it functions properly.
If required, adjust the sensitivity with a screwdriver at the right
side foot end of the retracting frame assembly.
Central brake and steer
Test the brake casters to determine if the bed moves when you
activate the brake mode.
Test the steer mode to determine if the foot end casters lock in the
steer mode.
Caster tires
Check the tires for cuts, wear, tread life, etc.
Power cord and plug
Inspect the power cord and plug for cuts, nicks, or breaks.
Head and foot panels
Check the aesthetics.
Overall appearance
Check the general aesthetics of the bed.
Touch up the paint where necessary.
Inspect the labels, and replace as necessary.
Electrical test
Test the bed for electrical leakage.
Ground resistance must be less than 0.20 ohms.
Leakage current must be less than 65 microamps for 115 volt
models, and less than 90 micro amps for 230 volt models.
Footboard control panel
Test all the bed function switches in the nurse control panel for
proper operation of the function and momentary operation.
Test all of the lockout switches individually for proper operation.
LED indicators
Check all of the LEDs on the footboard to ensure proper opera-
tion.
Bed scale
Check the accuracy of the bed scale with different weights.
Test for proper operation of the scale functions and adjustments.
ZoneAire® Sleep
Surface
Check the mattress and air bladders for punctures, cuts, or tears.
Inspect the air compressor unit.
Check all of the hose connections (to the mattress and inside the
mattress) and O-ring to prevent air leaks.
Check each zone for function.
Check all switches and the CPR release.
Function
Procedure
Summary of Contents for Advanta P1600
Page 2: ......
Page 6: ...Revisions Page iv Advanta Bed Service Manual man168ra NOTES...
Page 22: ...Chapter 1 Introduction Page 1 4 Advanta Bed Service Manual man168ra NOTES...
Page 142: ...Chapter 3 Theory of Operation Page 3 4 Advanta Bed Service Manual man168ra NOTES...
Page 310: ...Warranty Chapter 5 Parts List Page 5 4 Advanta Bed Service Manual man168ra NOTES...
Page 353: ...Footboard P1606 Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 47 5 NOTES...
Page 371: ...Power Module Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 65 5 NOTES...
Page 386: ...Chapter 6 General Procedures Page 6 2 Advanta Bed Service Manual man168ra NOTES...
Page 404: ...Chapter 7 Accessories Page 7 2 Advanta Bed Service Manual man168ra NOTES...
Page 412: ...Schematic Wiring Diagram Patient Switch Board P N 67208 FO 3 2 m168a103 Back to Chapter 3...
Page 413: ...Schematic Wiring Diagram J Box UTV Board P N 64935 FO 3 3 m168_107 Back to Chapter 3...
Page 417: ...Schematic Wiring Diagram Logic Board PN 63285 FO 3 7 m168_115 Back to Chapter 3...
Page 418: ...Schematic Wiring Diagram Power Board PN 63349 FO 3 8 m168_118 Back to Chapter 3...
Page 419: ...Schematic Wiring Diagram Frame Interface Board PN 63352 FO 3 9 m168_120 Back to Chapter 3...
Page 422: ...Schematic Wiring Diagram Sensor Board PN 4474001 FO 3 12 m168_128 Back to Chapter 3...
Page 425: ...Schematic Wiring Diagram Head Sensor Assembly P N 63888 FO 3 14 m168a186 Back to Chapter 3...
Page 427: ...Schematic Wiring Diagram HIlow Knee Sensor PN 63885 FO 3 16 m168_178 Back to Chapter 3...