ZoneAire® Sleep Surface Software Theory of Operation
Chapter 3: Theory of Operation
Page 3 - 44
Advanta™ Bed Service Manual (man168ra)
Continuous Run Error
In all modes, except the off mode, the system is monitored for the following
failure conditions that would cause the pump to run continuously:
• Excessive air leaks
• Faulty pump
• Faulty air manifold or flow direction valve
• Faulty pressure sensor or sensor circuit
• Faulty control circuit
• Loss of AC power or blown fuse
• Open or pinched hose
• Mattress connector not seated properly
If the pump runs continuously for approximately 10 min, in either an inflate or
deflate sequence, a continuous run error is flagged, the operating mode
transitions to off, and the footboard control panel service required LED
repeatedly flashes code 2, 2.
Triac Short Error
In all modes, except the off mode, the system is repeatedly checked for a
control circuit failure that would cause an air manifold valve, the flow
direction valve, or the pump to be wrongly activated.
Whenever the system is not correcting pressures, the feedbacks from the triac
drive circuits are monitored. If any feedback indicates an activated drive circuit
continuously for 5 seconds, a triac short error is flagged, the operating mode
transitions to off, and the footboard control panel service required LED
repeatedly flashes code 2, 3.
EEPROM Error
Upon reset and during power up (transition from the off mode to pressure relief
mode), the calibrations read from the pressure sensing circuit EEPROM are
checked for validity. Each calibration constant is stored in four locations to aid
in determining whether there is an error. The circuit also detects faulty
readings due to shorted or open communication lines between the control
circuit and the pressure sensing circuit.
Summary of Contents for Advanta P1600
Page 2: ......
Page 6: ...Revisions Page iv Advanta Bed Service Manual man168ra NOTES...
Page 22: ...Chapter 1 Introduction Page 1 4 Advanta Bed Service Manual man168ra NOTES...
Page 142: ...Chapter 3 Theory of Operation Page 3 4 Advanta Bed Service Manual man168ra NOTES...
Page 310: ...Warranty Chapter 5 Parts List Page 5 4 Advanta Bed Service Manual man168ra NOTES...
Page 353: ...Footboard P1606 Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 47 5 NOTES...
Page 371: ...Power Module Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 65 5 NOTES...
Page 386: ...Chapter 6 General Procedures Page 6 2 Advanta Bed Service Manual man168ra NOTES...
Page 404: ...Chapter 7 Accessories Page 7 2 Advanta Bed Service Manual man168ra NOTES...
Page 412: ...Schematic Wiring Diagram Patient Switch Board P N 67208 FO 3 2 m168a103 Back to Chapter 3...
Page 413: ...Schematic Wiring Diagram J Box UTV Board P N 64935 FO 3 3 m168_107 Back to Chapter 3...
Page 417: ...Schematic Wiring Diagram Logic Board PN 63285 FO 3 7 m168_115 Back to Chapter 3...
Page 418: ...Schematic Wiring Diagram Power Board PN 63349 FO 3 8 m168_118 Back to Chapter 3...
Page 419: ...Schematic Wiring Diagram Frame Interface Board PN 63352 FO 3 9 m168_120 Back to Chapter 3...
Page 422: ...Schematic Wiring Diagram Sensor Board PN 4474001 FO 3 12 m168_128 Back to Chapter 3...
Page 425: ...Schematic Wiring Diagram Head Sensor Assembly P N 63888 FO 3 14 m168a186 Back to Chapter 3...
Page 427: ...Schematic Wiring Diagram HIlow Knee Sensor PN 63885 FO 3 16 m168_178 Back to Chapter 3...