Preventive Maintenance
Chapter 6: General Procedures
Page 6 - 10
Advanta™ Bed Service Manual (man168ra)
Preventive Maintenance Schedule
Table 6-2. Preventive Maintenance Schedule
Function
Procedure
Head limits
Run the head section to the full upper and lower limits to ensure
proper function of the limit switches.
Knee limits
Run the knee section to the full upper and lower limits to ensure
proper function of the limit switches.
Hilow limits
Run the bed to the full upper and lower limits to ensure proper
function of both head and foot hilow drive limit switches.
Contour limit
Run the head section up to ensure that the knee section rises to
15° to form a contour position.
Trendelenburg limits
Run the bed into both the Trendelenburg and Reverse
Trendelenburg positions to check the limits.
Check the degree indicator for accuracy.
Head drive screw
Inspect, clean, and lubricate the head drive screw with red lith-
ium grease.
Inspect the sliding head cover on the upper frame assembly.
Inspect the clip engagement, torque cage, and lift nut. Ensure that
they are in proper working order.
Knee drive screw
Inspect, clean, and lubricate the knee drive screw with red lith-
ium grease.
Hilow head drive screw
Inspect, clean, and lubricate the hilow head drive screw with red
lithium grease.
Hilow foot drive screw
Inspect, clean, and lubricate the hilow foot drive screw with red
lithium grease.
Motor capacitors
Check the capacitors to ensure that they still work.
Siderail controls
Test the siderail switches for proper operation.
Check for momentary operation of the switches.
PPM (Bed exit)
Ensure the PPM system works properly in all modes and that it
places the appropriate call when activated.
Replace worn or defective sensors if needed.
Communications
Inspect and test the communication junction box.
Test the SideCom® Communication System features for proper
operation.
Inspect the communication cable including the male and female
pins in the plug.
Test the Nurse Call super capacitor for proper operation.
Summary of Contents for Advanta P1600
Page 2: ......
Page 6: ...Revisions Page iv Advanta Bed Service Manual man168ra NOTES...
Page 22: ...Chapter 1 Introduction Page 1 4 Advanta Bed Service Manual man168ra NOTES...
Page 142: ...Chapter 3 Theory of Operation Page 3 4 Advanta Bed Service Manual man168ra NOTES...
Page 310: ...Warranty Chapter 5 Parts List Page 5 4 Advanta Bed Service Manual man168ra NOTES...
Page 353: ...Footboard P1606 Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 47 5 NOTES...
Page 371: ...Power Module Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 65 5 NOTES...
Page 386: ...Chapter 6 General Procedures Page 6 2 Advanta Bed Service Manual man168ra NOTES...
Page 404: ...Chapter 7 Accessories Page 7 2 Advanta Bed Service Manual man168ra NOTES...
Page 412: ...Schematic Wiring Diagram Patient Switch Board P N 67208 FO 3 2 m168a103 Back to Chapter 3...
Page 413: ...Schematic Wiring Diagram J Box UTV Board P N 64935 FO 3 3 m168_107 Back to Chapter 3...
Page 417: ...Schematic Wiring Diagram Logic Board PN 63285 FO 3 7 m168_115 Back to Chapter 3...
Page 418: ...Schematic Wiring Diagram Power Board PN 63349 FO 3 8 m168_118 Back to Chapter 3...
Page 419: ...Schematic Wiring Diagram Frame Interface Board PN 63352 FO 3 9 m168_120 Back to Chapter 3...
Page 422: ...Schematic Wiring Diagram Sensor Board PN 4474001 FO 3 12 m168_128 Back to Chapter 3...
Page 425: ...Schematic Wiring Diagram Head Sensor Assembly P N 63888 FO 3 14 m168a186 Back to Chapter 3...
Page 427: ...Schematic Wiring Diagram HIlow Knee Sensor PN 63885 FO 3 16 m168_178 Back to Chapter 3...