ZoneAire® Sleep Surface Software Theory of Operation
Chapter 3: Theory of Operation
Page 3 - 36
Advanta™ Bed Service Manual (man168ra)
Communication between the controller circuit and the footboard control panel
is via LON communications protocol through a single pair of wires utilizing
RS485 medium standard. Communications between the controller circuit and
the pressure sensing circuit is via a 25 pin D-sub cable, utilizing 3-wire,
synchronous, serial, bi-directional communications (along with other discrete
lines, and AC and DC power).
General Description
The ZoneAire® Sleep Surface is an air-inflated, multi-functional, automatic
sleep surface used in conjunction with hospital patient beds. The system
functions are separated into several modes of operation.
The pressure relief mode is designed to minimize and uniformly distribute
interface pressures between a patient and the sleep surface.
Three heel relief modes further reduce heel-to-sleep surface interface pressure
and can accommodate a wide range of patient heights.
The auto firm mode inflates the mattress to a high pressure to aid in transfer of
patients to and from the bed.
The CPR mode inflates the sleep surface to a high pressure to provide a firm
surface should CPR need to be performed on a patient.
Software algorithms are included to provide limited self-diagnostics to reduce
the probability of unknowingly operating in a degraded state and/or reduce
durability of components. Software algorithms, which can be activated through
the footboard control panel, are also included to aid in troubleshooting.
Reset
A reset can happen two ways. A reset is initiated when power is applied to the
Neuron micro controller, and when a Neuron watchdog timer expires.
Summary of Contents for Advanta P1600
Page 2: ......
Page 6: ...Revisions Page iv Advanta Bed Service Manual man168ra NOTES...
Page 22: ...Chapter 1 Introduction Page 1 4 Advanta Bed Service Manual man168ra NOTES...
Page 142: ...Chapter 3 Theory of Operation Page 3 4 Advanta Bed Service Manual man168ra NOTES...
Page 310: ...Warranty Chapter 5 Parts List Page 5 4 Advanta Bed Service Manual man168ra NOTES...
Page 353: ...Footboard P1606 Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 47 5 NOTES...
Page 371: ...Power Module Chapter 5 Parts List Advanta Bed Service Manual man168ra Page 5 65 5 NOTES...
Page 386: ...Chapter 6 General Procedures Page 6 2 Advanta Bed Service Manual man168ra NOTES...
Page 404: ...Chapter 7 Accessories Page 7 2 Advanta Bed Service Manual man168ra NOTES...
Page 412: ...Schematic Wiring Diagram Patient Switch Board P N 67208 FO 3 2 m168a103 Back to Chapter 3...
Page 413: ...Schematic Wiring Diagram J Box UTV Board P N 64935 FO 3 3 m168_107 Back to Chapter 3...
Page 417: ...Schematic Wiring Diagram Logic Board PN 63285 FO 3 7 m168_115 Back to Chapter 3...
Page 418: ...Schematic Wiring Diagram Power Board PN 63349 FO 3 8 m168_118 Back to Chapter 3...
Page 419: ...Schematic Wiring Diagram Frame Interface Board PN 63352 FO 3 9 m168_120 Back to Chapter 3...
Page 422: ...Schematic Wiring Diagram Sensor Board PN 4474001 FO 3 12 m168_128 Back to Chapter 3...
Page 425: ...Schematic Wiring Diagram Head Sensor Assembly P N 63888 FO 3 14 m168a186 Back to Chapter 3...
Page 427: ...Schematic Wiring Diagram HIlow Knee Sensor PN 63885 FO 3 16 m168_178 Back to Chapter 3...