100
correction constant and the difference between the compensator
axis and the vertical axis of the instrument. Press ‘ENT’ to save
the new calculated parameters, otherwise press ‘ESC’ to quit and
the previous correction constant remains.
9.6.3 Calibrate TILT:Y
The steps of this function are same as them of calibration
Tilt X, but you must turn the instrument 90 degrees before every
pressing [Enter], if the reading is stable, press [Enter].
9.7 Config (Instrument constant)
Operations are permitted only when the instrument was
tested strictly. We recommend you to set it after it is tested by
the factory or professional verification institutions. Notice: the
multiplication constant is zero.
The input operation of additive constant (Add const) is
same as it of multiplication constant (Mul. const).Here, take
additive constant for example.
In the ‘Config’ menu interface, press [1] to enter the
interface of inputting additive constant, as shown in the picture
below:
B.S
Enter
Add C
6 mm
After inputting, press [Enter] with a prompt “Saved”
appearing, then exit the interface.