“Adjust”:
Adjust
1. Calibrate I.E
2. Calibrate TILT:X
3. Calibrate TILT:Y
9.6.1 Calibrate I.E
To adjust index error, the system will first ask you to ‘Aim
at a target F1’(face left). You may press [ENT] to confirm after
aiming, and the instrument will measure its vertical angle
automatically. The angle will be displayed for one second and
the system will ask you to ‘Aim at a target F2’(face right).
Following the same steps, the instrument will then calculate the
index error and display it.
Now press [ENT] to save the index error (calibration done),
otherwise press [ESC] to quit (previous I.E remains).
9.6.2 Calibrate TILT:X
When calibrating tilt, please make sure that the tilt is off
and the index error is little.
First,
p
lace the instrument as picture shown below with
collimator facing up. This will help screw C to adjust the
inclination of the instrument.
A
B
C
Enter the interface of tilt calibration: