Section 8-2 - Cover Replacement Procedures
8-2-13
Front Handle Replacement Procedure
8-2-13-1
Tools
Use the appropriate Phillips-type screw drivers and an 8mm wrench, to replace the
right
and
left
probe
holders.
8-2-13-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-2-13-3
Front Handle Removal Procedure
1) Remove the bottom keyboard cover, as described in
8-2-8-3 "Bottom Keyboard Cover Removal
Procedure" on page 8-14
.
2) Remove the two 8mm securing nuts (located below the keyboard) that fasten each of the probe
holders to the metal frame - refer to
Figure 8-24 "Probe Holder Nuts" on page 8-23
.
3) Remove the probe holders.
Figure 8-25 Front Handle Securing Nuts
4.) Remove the five 8mm securing nuts (located below the keyboard) that fasten Front Handle to the
metal frame, as shown in
Figure 8-25
, above.
5) Remove the Front Handle.
8-2-13-4
Front Handle Installation Procedure
1) Return the Front Handle to its original position on the front of the machine.
2) Fasten each side of the Front Handle beneath the keyboard with its 8 mm securing nuts (five in
total), as seen in
Figure 8-25
, above.
3) Return the probe holders to their original position on either side of the keyboard.
4) Fasten each probe holder beneath the keyboard with the two (8 mm) securing nuts, shown in
Figure
8-24 "Probe Holder Nuts" on page 8-23
.
5) Return the bottom keyboard cover to its original position, as described in
8-2-8-4 "Bottom Keyboard
Cover Installation Procedure" on page 8-15
.
.