Chapter 2 Pre-Installation
2-3
2-2-5
Electrical Requirements
NOTE:
GE Medical Systems requires a dedicated power and Ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-5-1
Vivid 3N Pro/Expert Power Requirements
Electrical specifications for the Vivid 3N Pro/Expert ultrasound scanner, monitor and onboard
peripherals are as follows:
2-2-5-2
Inrush Current
Inrush current is not a factor for consideration, due to the inrush current limiting properties of the power
supplies.
Maximum power requirement = 1.2 KVa
•
100V AC: 8A
•
120V AC: 8A
•
220 - 240V AC: 4A
2-2-5-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
Table 2-10 Electrical Requirements, All Models
Voltage
Tolerances
Op. Current
Frequency
120 VAC
±10%
8A
50-60 Hz
230 VAC
±10%
4A
50-60 Hz
CAUTION
POWER OUTAGE MAY OCCUR.
The Vivid 3N Pro/Expert requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.