Chapter 3 Installation
3-33
3-6-5-3
Disconnecting the Ultrasound Unit from the Electrical Outlet
CAUTION: Whenever disconnecting the Vivid 3N unit from the electrical outlet, always observe
the safety precautions. First unplug the mains power cable from the wall outlet socket, then from
the unit itself. Remove by pulling on the cable connector - DO NOT pull on the cable.
1) Turn OFF the AC circuit breaker on the rear of the unit.
2) Unplug the mains power cable from the AC wall outlet socket.
3)
Unplug the mains power cable connector from the AC input socket (refer to
Figure 3-17 "Circuit
Breaker and Power Cable on Back of Scanner" on page 3-32
).
3-6-6
Switching the System ON/OFF
3-6-6-1
Switching the System ON
1) Verify that the ultrasound unit has been connected to the power supply and that the circuit breaker
is ON, as described in
3-6-5-2 "Connecting the Ultrasound Unit to the Electrical Outlet" on page 3-
32
.
2) Hold down the
On/Off
button on the control panel for 3 seconds. The system automatically
performs an initialization sequence which includes the following:
•
Loading the operating system.
•
Running a quick diagnostic check of the system.
•
Detecting connected probes.
The system first enters 2D-Mode with the probe and application that were last used before the system
was shut down. If the probe has been removed since the system was last used, the currently connected
probes and their available applications are displayed and selected by default.
NOTICE
Disconnecting the Mains Power Cable
before
switching OFF the Circuit Breaker will activate the
uninterruptible power system (UPS) in the back-end processor, forcing an ordered shutdown of the
system.
NOTICE
When AC power is applied to the scanner, the
On/Off
button on the control console illuminates amber,
indicating the Back-end Processor is in
Standby
mode.