Chapter 8 Replacement Procedures
8-15
8-2-8-3 Bottom Keyboard Cover Removal Procedure
(cont’d)
4) Lower the rear end of the bottom keyboard cover and free it from the Up/Down handle by passing
the handle through the slot on the cover, as shown
Figure 8-16
below.
Figure 8-16 Removing the Bottom Keyboard Cover
8-2-8-4
Bottom Keyboard Cover Installation Procedure
1) Raise the control console monitor to its
maximum
height.
2) Position the bottom keyboard cover so that the Up/Down (height adjustment) handle is correctly
aligned with the slot on the cover.
3) With one hand, lift the bottom keyboard cover upwards while using the other hand to guide the
Up/Down handle down through the slot in the cover. Push the cover all the way up to return it to its
original location (refer to
Figure 8-16
, above).
4) Return the cable hooks to their original position underneath the keyboard and fasten them securely
with their screws - refer to
Figure 8-15 "Removing Probe Cable Hooks under Keyboard Cover" on
page 8-14
.