Chapter 3 LOGIQ E9 Setup
3 - 19
3-6-3
Electrical specifications
3-6-3-1
Verification of the
LOGIQ E9
’s voltage setting
Verify that the mains voltage specified for the unit is available on-site.
The voltage setting for the unit is found on a label on the back of the LOGIQ E9 on lower rear frame of
the LOGIQ E9.
3-6-3-2
Electrical specifications for LOGIQ E9
Table 2-3 "Electrical specifications for the LOGIQ E9" on page 2-3
for the electrical specifications
for LOGIQ E9 include LCD and on board peripherals.
3-6-4
Connections on the I/O Rear Panel
NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according
to the respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-
1 for medical equipment). Furthermore, all complete configurations shall comply with the valid
version of the system standard IEC60601-1-1. Everybody who connects additional equipment
to the signal input part or signal output part of LOGIQ E9, configures a medical system, and is
therefore responsible that the system complies with the requirements of the valid version of
IEC60601-1-1. If in doubt, consult the technical service department or your local representative
for GE.
WARNING
WARNING
CONNECTING A LOGIQ E9 UNIT TO THE WRONG VOLTAGE LEVEL WILL MOST
LIKELY DESTROY THE UNIT.